We suggest that your practice establish an internal calibration program for your laser. Recalibration is
recommended at least once per year based on typical usage.
Recalibration may be performed by Den-Mat
®
Holdings, LLC by returning the unit. As an alternative, you may
purchase a calibrated hand-held power meter approved for use with 808 nm devices to check power output.
The laser should be set at 1, 2, and 3W in CW with output checked at each level. Similarly, the output should be
set 1, 3, and 5W in P with output checked at each level. The output display on the hand-held meter should be
within 20% of each setting. If not, replace the Fiber Optic Tip and recheck power outputs. If output displays are
still outside the ±20% tolerance, return the unit to Den-Mat Holdings, LLC for recalibration. There are no methods
available for the user to adjust the calibration of the unit and the Handpiece must not be opened by the user for
any reason.